PULSE 6 SERVICE MANUAL
PAGE 7.17
SEPT 2011SUNRISE MEDICAL
Section 7
Caution: the actuator and micro-switch assembly wires are fragile. Use caution when removing
the actuator and micro-switch assembly.
8. Remove the actuator and micro-switch assembly
as one part if the entire assembly is to be replaced
(fi gure 7.17.1). Reverse the procedures to install
and retest as necessary.
9. Note the connections of the wires and separate
them (fi gure 7.17.2). Remove the actuator.
Removal Procedures (cont.)
fig. 7.17.1
fig. 7.17.2
Note: Perform this step if only the actuator is to be replaced.
10. Reverse the previous procedure to install
the actuator/micro-switch assembly. Retest as
necessary.