Quickie Pulse Wartungshandbuch Seite 32

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 72
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 31
PULSE 6 SERVICE MANUAL
PAGE 4.1
SEPT 2011SUNRISE MEDICAL
Section 4
R-net Troubleshooting Procedures
Chair Will Not Power Up
2
1
3
1. Check for battery voltage at the hand control
using a Multimeter. Connect the Multimeter
between the two outside pins (figure 4.1.1), pin 1
is positive (+), pin 2 is negative (-). Note: Positive
is on the right.
2. If battery voltage is present, replace the
following components in this order:
a. Cables
b. Hand control
c. Control module. (refer to the Control Module
Replacement section of this manual.)
d. Retest as necessary.
3. Verify that the buss cables are correctly mated
between the hand control and the control module
(figure 4.1.2).
Note: This connector is mated incorrectly.
The connectors are designed to visually
indicate when they are not mated correctly.
If yellow is showing between the halves (A),
push them closer together until only black is
visible. Retest as necessary.
A
Fig. 4.1.1
Fig. 4.1.2
Seitenansicht 31
1 2 ... 27 28 29 30 31 32 33 34 35 36 37 ... 71 72

Kommentare zu diesen Handbüchern

Keine Kommentare