PULSE 6 SERVICE MANUAL
PAGE 0.16
SEPT 2011 SUNRISE MEDICAL
R-NET Plugs/Connectors
M1 = LEFT SIDE MOTOR
M2 = RIGHT SIDE MOTOR
INH-2 = INHIBIT 2
A1 = ACTUATOR 1
A2 =ACTUATOR 2
OBC = ON BOARD
CHARGER
+ - =BATTERY
1 = 24 Vdc (POS)
2 = 0 Vdc
3 = Inhibit (NEG)
1 = White (-)
2 = Blue
3 = Red (+)
4 = Black (-)
2
1
3
Motor Plug Port
1 = 24 Vdc
2 = 0 Vdc
3 = INHIBIT 1/
PROGRAMMER
Rnet Controller
12
3
4
_
+
+
_
Charger Port
On-Board Charger - Used to power Switch-
controlled actuators and Tilt Inhibit.