PULSE 6 SERVICE MANUAL
PAGE 5.9
SEPT 2011SUNRISE MEDICAL
Section 5
1 to 3 ohms
Positive probe here
Negative probe here
Continuity between pins
Positive probe here
Negative probe here
Open between pins
Negative probe here
Positive probe here
Open between pins
Negative probe here
Positive probe here
Continuity between pins
Note: Since the tilt actuator will not operate,
it is only possible to check for one condition.
Pick the condition below that best describes
the position of the tilt system.
9. Example 1, the tilt is in mid-stroke. The value
recorded is the resistance through the actuator
motor windings (Figure 5.9.1). If this reading is
not correct, replace the tilt actuator and retest.
Note: The value recorded in one direction is
the forward resistance through a diode and
will vary with the type of Multimeter used.
This value is not important, as long as there
is continuity in one direction only. This value
may even be high. The value recorded in the
other direction is across the open contact of
the end of stroke micro-switch and should be
open.
10. Example 2, the tilt is in the full down position
(Figure 5.9.2). If the readings are not correct,
replace the tilt actuator. Refer to Tilt Actuator
and Micro-Switch Removal section this manual.
Retest as necessary.
Note: The value recorded in one direction is
the forward resistance through a diode and
will vary with the type of Multimeter used.
This value is not important, as long as there
is continuity in one direction only. This value
may even be high. The value recorded in the
other direction is across the open contact of
the end of stroke micro-switch and should be
open.
11. Example 3, the tilt is in the full up position
(fi gure 5.9.3). If the readings are not correct,
replace the tilt actuator. Refer to Tilt Actuator
and Micro-Switch Removal section this manual.
Retest as necessary.
*SEE NOTE*
R-net Fault Codes cont.
fig. 5.9.1
fig. 5.9.2
fig. 5.9.3
*SEE NOTE*