PULSE 6 SERVICE MANUAL
PAGE 4.2
SEPT 2011 SUNRISE MEDICAL
Section 4
R-net Troubleshooting Procedures (cont.)
5. Disconnect the power connector from the
control module, and use a Multimeter to check
for battery voltage at the connector (figure 4.2.1).
If voltage is present, replace the control module.
Retest as necessary.
Note: The power connector is the larger 2-pin
connector between the left and right motor
cable.
6. Manually tilt the seat back and remove the
shroud from the base. Refer to Battery Removal
section of this manual.
fig 4.2.1
7. For chairs prior to PLS-100808, verify that
the circuit breaker located at the front of the
power chair is not tripped (figure 4.2.2). Reset if
it is tripped and retest.
fig. 4.2.2
8. Disconnect the batteries. (also see Sub-
section 7.1)