10©2006 Sunrise MedicalINTRODUCTIONHealth and SafetyWhile working on powered mobility products, it is essential to observe good working practices. Bel

Seite 3 - Table of Contents

100©2006 Sunrise MedicalSEATING SECTION Lift SystemEleven inches of lift. • Linak Actuator.• Cross arm x-style actuation design.• Connects directly to

Seite 4 - Introduction

101©2006 Sunrise MedicalSEATING SECTIONSeat InterfacePermanent Tilt AdjustmentsThe Perfect Fit Seating System may be mounted with 0, 3, 6 or 9 degrees

Seite 5 - QUICKIE GROOVE FWD

102©2006 Sunrise MedicalSEATING SECTIONPower Recline with Power Shear ReductionThe New Power Recline with Power Shear Reduction is modeled after the Q

Seite 6 - Diagnostics Chart

103©2006 Sunrise MedicalSEATING SECTIONLocate the Recline Counter Reset Switch (CRS) 1. mounted on the left seat hinge (Fig 9.90). Note: this switch

Seite 7 - ©2002 Sunrise Medical

104©2006 Sunrise MedicalSEATING SECTIONConnect a meter to the 2 pins indicated in the 5. photo (Fig 3.93). Check continuity through the switch by pre

Seite 8 - MULTIMETER

105©2006 Sunrise MedicalSEATING SECTIONAdditionally there may be encoder errors. Use the HHP to re-home the chair prior to completing any tests. Com

Seite 9

106©2006 Sunrise MedicalSEATING SECTIONConnect a meter to the 2 pins indicated in the 5. photo (Fig 3.99). Check continuity through the switch by pre

Seite 10 - 10©2006 Sunrise Medical

107©2006 Sunrise MedicalSEATING SECTIONThe Recline Actuator will not run, or runs erratically Complete the following procedures to test the Recline A

Seite 11 - ©2006 Sunrise Medical

108©2006 Sunrise MedicalSEATING SECTIONFollow the cable to another connector leading to 5. the actuator if the resistance is excessive. Disconnect th

Seite 12 - 12©2006 Sunrise Medical

109©2006 Sunrise MedicalSEATING SECTIONIf the resistance is open, follow the cable to an-5. other cable leading to the actuator. Disconnect this conn

Seite 13

11©2006 Sunrise MedicalINTRODUCTIONRemove personal items of jewelry, such as rings, watches, chains etc. • before working on batteries. Such items cou

Seite 14 - INTRODUCTION

110©2006 Sunrise MedicalSEATING SECTIONAdjustments of the New Recline for the seat-pan are similar to the old Perfect Fit Recline. Refer to that por-

Seite 15 - Discharge

111©2006 Sunrise MedicalSEATING SECTIONSeat-back depth is adjusted by moving the mount-7. ing brackets to the front or rear in the seat-pan track. Lo

Seite 16 - 16©2006 Sunrise Medical

112©2006 Sunrise MedicalSEATING SECTIONIn order to remove either the recline or the shear actua-tor, the gas strut must be removed. In addition, the Q

Seite 17

113©2006 Sunrise MedicalSEATING SECTIONOnce both screws are loosened, lift the QR-MAC 5. mount up and place it to the side (Fig 3.18).WARNINGDuring th

Seite 18 - 18©2006 Sunrise Medical

114©2006 Sunrise MedicalSEATING SECTIONUse a ¼” Hex wrench and a 9/16” open-end 12. wrench and remove the top bolt that holds the Re-cline actuator to

Seite 19

115©2006 Sunrise MedicalSEATING SECTIONShear Actuator RemovalIn order to remove the Shear Actuator, the Gas Strut and Recline Actuator must be removed

Seite 20 - Fig 1.10

116©2006 Sunrise MedicalSEATING SECTIONIn order to eliminate the re-home switch (previously found on the Smart Seat), a separate Counter Reset Switch

Seite 21

117©2006 Sunrise MedicalSEATING SECTIONIn order to eliminate the re-home switch (previously found on the Smart Seat), a separate Counter Reset Switch

Seite 22 - Fig 1.17

118©2006 Sunrise MedicalSEATING SECTIONAcronyms for Delphi APPENDIX AAPPENDIX AAAcronym or Term Defi nitionACM Attendant Control ModuleBSOC Battery

Seite 23

2009 Quickie Rhythm Service Manual Supplement©2009 Sunrise Medical Inc. xxxxxx Rev APG DRIVES TECHNOLOGYR

Seite 24 - 24©2006 Sunrise Medical

12©2006 Sunrise MedicalINTRODUCTIONSources of EMI 1. Hand-Held Transceivers: The antenna is usually mounted directly on the unit.These include:Citizen

Seite 25 - Fig 1.32

120©2006 Sunrise MedicalSEATING SECTIONQuickie Rhythm Service Manual ContentsVR2 Remote Controller ...0.3VR2 Plugs/Con

Seite 26 - Fig 1.36

121©2006 Sunrise MedicalSEATING SECTION RHYTHM SERVICE MANUAL2009PAGE 0.3SUNRISE MEDICALVR2 Remote ControllerVR2 Controller ButtonsBattery Gauge A se

Seite 27 - Fig 1.39

RHYTHM SERVICE MANUALPAGE 0.42009SUNRISE MEDICALVR2 Plugs/Connectors1 = 24 Vdc2 = 0 Vdc3 = Inhibit 1/ Programmer1 = Black (-)2 = White3 = Yellow 4 =

Seite 28 - Fig 1.41

RHYTHM SERVICE MANUAL2009PAGE 0.5SUNRISE MEDICALVR2 -Main Wiring DiagramsDriveRightDriveLeftRedBlack2-waySB50RedBattery2-waySB50BlackRedBlackBatteryF

Seite 29 - Fig 1.46

RHYTHM SERVICE MANUALPAGE 0.62009SUNRISE MEDICALDriveRightDriveLeftRedBlack2-waySB50RedBattery2-waySB50BlackRedBlackBatteryFuseDual-ToggleSingle Actu

Seite 30 - Fig 1.49

RHYTHM SERVICE MANUAL2009PAGE 0.7SUNRISE MEDICALR-NET Remote ControllerOn/Off Key- Press to power on or off the power chair or Controller.Speed/ Pr

Seite 31 - Fig 1.53

RHYTHM SERVICE MANUALPAGE 0.82009SUNRISE MEDICALR-NET Remote ControllerOn/Off Key- Press to power on or off the power chair or Controller.Speed/ Pr

Seite 32 - 70mm Power options

RHYTHM SERVICE MANUAL2009PAGE 0.9SUNRISE MEDICALR-NET Optional ElectronicsR-net OmniR-net Output ModuleR-net Power ModuleR-net Blue Tooth ModuleR-net

Seite 33 - Fig 1.59

RHYTHM SERVICE MANUALPAGE 0.102009SUNRISE MEDICALR-NET Plugs/ConnectorsM1 = RIGHT SIDE MOTORM2 = LEFT SIDE MOTORJSM = JOYSTICK MODULEINH-2 = INHIBIT

Seite 34 - Fig 1.62

RHYTHM SERVICE MANUAL2009PAGE 0.11SUNRISE MEDICALR-NET Plugs/Connectors1 2 3 4 56 7 8 9On/Off InputSip and Puff PortUser Jack 2User Jack 1Port 2 &ap

Seite 35 - Fig 1.63

13©2006 Sunrise MedicalINTRODUCTIONTo help prevent Electro Static Discharge (ESD) the following proper handling techniques should be followed:ESD:Do n

Seite 36 - 36©2006 Sunrise Medical

RHYTHM SERVICE MANUALPAGE 0.122009SUNRISE MEDICALMain Wiring Diagram R-NETR-NET DRIVE ONLYDriveRightDriveLeftRedBlack2-waySB50RedBattery2-waySB50Blac

Seite 37 - Fig 1.69

RHYTHM SERVICE MANUAL2009PAGE 0.13SUNRISE MEDICALMain Wiring Diagram R-netDR IVE-THR UHARNE SSR-NET DRIVE-THRU TILTDriveRightDriveLeftRedBlack2-waySB

Seite 38 - 38©2006 Sunrise Medical

RHYTHM SERVICE MANUALPAGE 0.142009SUNRISE MEDICAL

Seite 39 - Quickie Electronics by Delphi

RHYTHM SERVICE MANUALPAGE 1.12009SUNRISE MEDICALSection 1Battery Connection Test1. Check that the female VR2 Bus plug on the chair has voltage. Set

Seite 40 - Fig 2.4.2

RHYTHM SERVICE MANUAL2009PAGE 1.2SUNRISE MEDICALTroubleshooting: No Power (cont.)Check Battery Wire HarnessCheck that the battery wire harness has th

Seite 41 - Fig 2.4.3

RHYTHM SERVICE MANUALPAGE 2.12009SUNRISE MEDICALSection 2VR2 Remote Controller DisplayIndicates that the wheelchair is locked. To unlock the wheelch

Seite 42 - 42©2006 Sunrise Medical

RHYTHM SERVICE MANUAL2009PAGE 2.2SUNRISE MEDICALNote: On hand controls that contain LED battery gauges, they will display the fault code byilluminat

Seite 43

y k RHYTHM SERVICE MANUALPAGE 2.32009SUNRISE MEDICALE-StopRestartWhen the control system requires a reboot this system will be flashing.Recycle power

Seite 44 - Fig 2.10.2

RHYTHM SERVICE MANUAL2009PAGE 2.4SUNRISE MEDICALSection 2

Seite 45 - Fig 2.12

RHYTHM SERVICE MANUALPAGE 3.12009SUNRISE MEDICALSection 3Check that the batteries are fully charged and in good condition; and check all cables and c

Seite 46 - 46©2006 Sunrise Medical

14©2006 Sunrise MedicalINTRODUCTION

Seite 47

RHYTHM SERVICE MANUAL2009PAGE 3.2SUNRISE MEDICALSection 3Check that the batteries are fully charged and in good condition; and check all cables and c

Seite 48 - 48©2006 Sunrise Medical

RHYTHM SERVICE MANUALPAGE 3.32009SUNRISE MEDICALSection 3VR2 Controller Diagnostics Codes (cont.)Check that the batteries are fully charged and in go

Seite 49 - Fig 2.17

RHYTHM SERVICE MANUAL2009PAGE 3.4SUNRISE MEDICALSection 3VR2 Controller Diagnostics Codes (cont.)Check that the batteries are fully charged and in go

Seite 50 - Fig 2.18

RHYTHM SERVICE MANUALPAGE 3.52009SUNRISE MEDICALSection 3VR2 Controller Diagnostics Codes (cont.)Nine Bars - Solenoid Brake TripBattery FaultCheck th

Seite 51 - Fig 2.20

RHYTHM SERVICE MANUAL2009PAGE 3.6SUNRISE MEDICAL

Seite 52 - Fig 2.22

RHYTHM SERVICE MANUALPAGE 4.12009SUNRISE MEDICALSection 4R-net Troubleshooting ProceduresChair Will Not Power Up2131. Check for battery voltage at t

Seite 53 - MotorsHarnessBattery

RHYTHM SERVICE MANUALPAGE 4.22009SUNRISE MEDICALSection 45. Disconnect the power connector from the control module, and use a Multimeter to check fo

Seite 54 - 54©2006 Sunrise Medical

RHYTHM SERVICE MANUALPAGE 4.32009SUNRISE MEDICALSection 410. If battery voltage is not present, use a Multimeter and measure for continuity across t

Seite 55 - QUICKIE ELECTRONICS SECTION

RHYTHM SERVICE MANUAL2009PAGE 4.4SUNRISE MEDICAL

Seite 56 - 56©2006 Sunrise Medical

RHYTHM SERVICE MANUALPAGE 5.12009SUNRISE MEDICALPower Chair Displays a Fault on the Hand Control or OmniThe R-Net control used on this power chair is

Seite 57

15©2006 Sunrise MedicalPOWER BASE SECTIONRhythm & Groove Power BasesBatteriesSafety If mishandled batteries can be dangerous and hazardous.All mob

Seite 58 - 58©2006 Sunrise Medical

RHYTHM SERVICE MANUAL2009PAGE 5.2SUNRISE MEDICALSection 5M1 Motor Error Verify that the left motor is connected to the control module.Check Motor res

Seite 59

RHYTHM SERVICE MANUALPAGE 5.32009SUNRISE MEDICALR-net Fault Codes cont.Section 5Bad Cable Inspect and replace the defective cable(s).Bad Settings Ver

Seite 60 - 60©2006 Sunrise Medical

RHYTHM SERVICE MANUAL2009PAGE 5.4SUNRISE MEDICALService for assistance. 1. Locate the 6-pin connector (A) at the rear of the tilt (fi g. 5.4.1). Mak

Seite 61 - Fig 2.23.15

r f RHYTHM SERVICE MANUALPAGE 5.52009SUNRISE MEDICAL Open Troubleshooting Tilt cont.Section 5Power Chair Will Not Drive in Creep Speed when Tilte

Seite 62 - Lt +0 Rt

RHYTHM SERVICE MANUAL2009PAGE 5.6SUNRISE MEDICALTilt Will Not Operate1. Selected the tilt mode on the Joystick or Omni. (fi g. 5.6.1). Operate the j

Seite 63

rRHYTHM SERVICE MANUALPAGE 5.72009SUNRISE MEDICALR-net Fault Codes cont.Section 5Warning: Use caution in the next step. Do not short the leads of t

Seite 64 - 64©2006 Sunrise Medical

RHYTHM SERVICE MANUAL2009PAGE 5.8SUNRISE MEDICAL 1 to 3 ohms Positive probe hereNegative probe hereContinuity between pinsPositive probe hereNegati

Seite 65 - Fig 2.23.21

RHYTHM SERVICE MANUALPAGE 6.12009SUNRISE MEDICALSection 5Troubleshooting Recline Power Chair Will Not Drive at Full SpeedPower Recline is offered on

Seite 66 - 66©2006 Sunrise Medical

RHYTHM SERVICE MANUAL2009PAGE 6.2SUNRISE MEDICALfig. 6.2.1Troubleshooting Power Recline Calibration.LeftLightsInhibit4Inhibit5RightLights123456Actuat

Seite 67

RHYTHM SERVICE MANUALPAGE 6.32009SUNRISE MEDICAL4. If you have a correct reading at this location, skip to step 6.If your reading is open, take a re

Seite 68 - 68©2006 Sunrise Medical

16©2006 Sunrise MedicalPOWER BASE SECTIONHow Long Should Batteries Last?An average of 1 to 1.5 years.Factors that affect battery performance:Maintenan

Seite 69

RHYTHM SERVICE MANUAL2009PAGE 6.4SUNRISE MEDICALSection 6Power Recline does not Operate1. Verify that the recline has not been disabled in programmi

Seite 70 - 70©2006 Sunrise Medical

RHYTHM SERVICE MANUALPAGE 6.52009SUNRISE MEDICALSection 6Power Shear Does not Operate1. To test the Shear actuator resistance, locate the ISM mounted

Seite 71 - Perfect Fit Seating System

RHYTHM SERVICE MANUAL2009PAGE 6.6SUNRISE MEDICAL

Seite 72 - 72©2006 Sunrise Medical

RHYTHM SERVICE MANUALPAGE 7.12009SUNRISE MEDICALTroubleshooting Power Legrest We are offering two versions of power legrest, Center Mount legrest act

Seite 73 - Fig 3.10 Fig 3.11

RHYTHM SERVICE MANUAL2009PAGE 7.2SUNRISE MEDICAL1 Are you able to access the legrest option on the joystick menu? If yes, contact Sunrise Medical T

Seite 74 - Fig 3.14

RHYTHM SERVICE MANUALPAGE 8.12009SUNRISE MEDICALSeating System Mounting InstructionsSection 8Non Tilt Module/Packer Interface

Seite 75 - Seat Pan

RHYTHM SERVICE MANUAL2009PAGE 8.2SUNRISE MEDICALASAP (Tilt Only) Seating System Mounting InstructionsSeat to Tilt MountingSection 8

Seite 76 - Fig 3.20

RHYTHM SERVICE MANUALPAGE 8.32009SUNRISE MEDICALTraditional Rehab Seating System Mounting InstructionsSeat to Tilt MountingSection 8

Seite 77

RHYTHM SERVICE MANUAL2009PAGE 8.4SUNRISE MEDICALTilt to Interface MountingSection 8

Seite 78 - Fig 3.25

RHYTHM SERVICE MANUALPAGE 8.52009SUNRISE MEDICALNew Power Recline Seating System Mounting InstructionsSeat to Recline MountingSection 8

Seite 79 - Fig 3.29

17©2006 Sunrise MedicalPOWER BASE SECTIONBattery Size Batteries function as a power wheelchair’s fuel • tank. The larger the group size, the farther t

Seite 80 - Fig 3.32

RHYTHM SERVICE MANUAL2009PAGE 8.6SUNRISE MEDICALfig. 7.7New Power Tilt & Recline Seating System Mounting InstructionsSeat to Tilt/Recline Mountin

Seite 81 - Fig 3.35

RHYTHM SERVICE MANUALPAGE 8.72009SUNRISE MEDICALfig. 7.11Seat to Floor Height MatrixActual STFH dimensions are listed for short confi guration:Add 1”

Seite 82 - 82©2006 Sunrise Medical

RHYTHM SERVICE MANUAL2009PAGE 8.8SUNRISE MEDICALSeat to Floor Height Matrix (cont) Actual STFH (in) Post Position Short Post Position Med Post Po

Seite 83

RHYTHM SERVICE MANUALPAGE2009SUNRISE MEDICALSeat to Floor Height Matrix (cont)T raditional R ehab Seating 15-24” W idth No Module 16.8 1 / A 3 / C

Seite 84 - Fig 3.45

RHYTHM SERVICE MANUAL2009PAGE 8.10SUNRISE MEDICALSection 8Asap Fixed Center mount Footrest Mounting Matrix Part # 014945 Position Part # 014941 Pos

Seite 85 - Fig 3.47

RHYTHM SERVICE MANUALPAGE 8.112009SUNRISE MEDICALSection 8Adjusting Power ELR / ALR Legrest LengthTo adjust & or remove the footrest length on a

Seite 86 - Fig 3.48

RHYTHM SERVICE MANUAL2009PAGE 8.12SUNRISE MEDICALSection 8Use a 13 mm open wrench and a 13mm socket to release the top nut. Fig 8.8Removing the legre

Seite 87 - Fig 3.49

RHYTHM SERVICE MANUALPAGE 8.132009SUNRISE MEDICALSection 8Adjusting Power Recline BackrestAdjustments of the New Recline for seat pan are similar to

Seite 88 - 88©2006 Sunrise Medical

RHYTHM SERVICE MANUAL2009PAGE 8.14SUNRISE MEDICALSection 8Removal Procedures (cont.)6. Seat depth is adjusted by moving the mounting brackets to the

Seite 89

RHYTHM SERVICE MANUALPAGE 9.12009SUNRISE MEDICALSection 9OBP Programming Quick Guide

Seite 90 - 90©2006 Sunrise Medical

18©2006 Sunrise MedicalPOWER BASE SECTIONGeneral Battery MaintenanceKeep terminals free of corrosion and insure wiring connections are secure. Check f

Seite 91 - Armature Pins

RHYTHM SERVICE MANUAL2009PAGE 9.2SUNRISE MEDICALSection 9OBP Programming Quick Guide

Seite 93 - Fig 3.66

182©2006 Sunrise Medical

Seite 94 - Fig 3.69

19©2006 Sunrise MedicalPOWER BASE SECTIONRhythmFirmly grasp the red battery connectors and pull 1. apart.Firmly grasp the black battery connectors and

Seite 95 - Fig 3.72

2©2006 Sunrise Medical

Seite 96

20©2006 Sunrise MedicalPOWER BASE SECTIONBattery ConnectorsInspecting the FusesRemove caps on each of the two 15 amp fuses and visually inspect. Insur

Seite 97 - SEATING SECTION

21©2006 Sunrise MedicalPOWER BASE SECTIONDesktop 8 Amp Fully Automatic ChargerPin 1 PositivePin 2NegativeDC Output ConnectorRed Light Green Light Func

Seite 98 - 98©2006 Sunrise Medical

22©2006 Sunrise MedicalPOWER BASE SECTIONSymptom Possible cause SolutionLED’s do not turn onNo Power.Unit not plugged in. Bad Outlet.Blown Fuse inter

Seite 99

23©2006 Sunrise MedicalPOWER BASE SECTIONMotorsBrake SolenoidMotorMotor ConnectorFig 1.19Fig 1.20Motor Motor ResistanceBrake Solenoid ResistanceStanda

Seite 100 - 100©2006 Sunrise Medical

24©2006 Sunrise MedicalPOWER BASE SECTIONOutput ShaftHubs and BoltsInspect: Tapered Shaft – Check keyway for wear – check for physical deteriorationUs

Seite 101 - Fig 3.89

25©2006 Sunrise MedicalPOWER BASE SECTION1. Remove Cotter Pin that holds the Large Freewheel release lever to the freewheel release on the gearbox.You

Seite 102 - 102©2006 Sunrise Medical

26©2006 Sunrise MedicalPOWER BASE SECTIONMotor Removal and Replacement GrooveSupport the Groove power base securely using jackstand or suitable blocki

Seite 103 - Fig 3.92

27©2006 Sunrise MedicalPOWER BASE SECTIONRemove the Bridging Plate by removing the four nuts, bolts and washers that attach it to the motor mount asse

Seite 104 - Fig 3.93

28©2006 Sunrise MedicalPOWER BASE SECTIONGroove Free Wheel AssemblyEngagedPush in - Rotate downFree WheelingFig 1.40Fig 1.41Fig 1.4212”1 1/4”3/8”5 3/4

Seite 105 - Fig 3.98

29©2006 Sunrise MedicalPOWER BASE SECTIONDisassemble:Carefully support the frame of the wheelchair us-1. ing jack stands.Remove the four lug nuts usin

Seite 106 - Fig 3.100

3 ©2006 Sunrise MedicalTable of ContentsDiagnostics Overview ...51Quickie by Delphi Diagnostic Codes .

Seite 107 - Fig 3.101

30©2006 Sunrise MedicalPOWER BASE SECTIONIntroduction Encoders allow chairs maintain a straight line while driving on uneven terrain and allows the us

Seite 108 - Fig 3.106

31©2006 Sunrise MedicalPOWER BASE SECTIONFinal Step is to Plug in ProgrammerGo to Program (Fig 1.50) Go to Motor control (Fig 1.51)Go into General Par

Seite 109 - Fig 3.108

32©2006 Sunrise MedicalPOWER BASE SECTIONSuspensionThe Quickie Rhythm base has a special 4-bar suspension system that pulls the front caster wheels up

Seite 110 - Fig 3.111

33©2006 Sunrise MedicalPOWER BASE SECTIONIt is important that you check the settings of the rear shock dampening assembly before you proceed to make a

Seite 111

34©2006 Sunrise MedicalPOWER BASE SECTIONFor Rhythm, each rear suspension element controls the pitch damper mechanism on that side of the chair. Timin

Seite 112 - Fig 3.117

35©2006 Sunrise MedicalPOWER BASE SECTIONAdjusting the timing of the anti-pitch effect is • accomplished by rotating the nut at the end of the anti-pi

Seite 113 - Fig 3.121

36©2006 Sunrise MedicalPOWER BASE SECTIONAnti-pitch Damper Mechanism CleaningIf the Anti-Pitch Mechanism becomes contaminated with grease or oil it wi

Seite 114 - 114©2006 Sunrise Medical

37©2006 Sunrise MedicalPOWER BASE SECTIONUsing a soft hammer tap down on the top of the cylin-der to release the internal parts.Use soap and water, al

Seite 115 - Fig 3.126

38©2006 Sunrise MedicalPOWER BASE SECTIONReplacing ShroudsRhythmThe Front (B) and Side shrouds (A) are held • in place by reusable shroud fasteners (C

Seite 116 - Fig 3.129

39©2006 Sunrise MedicalQUICKIE ELECTRONICS SECTIONQuickie Electronics by DelphiHand Control Types and FunctionalityActuator ModesPower – On/OffHorn Sp

Seite 117 - Fig 3.131

4©2006 Sunrise MedicalIntroductionPlease read and follow instructions in this service manual before attempting to troubleshoot or repair this product

Seite 118 - Acronyms for Delphi

40©2006 Sunrise MedicalQUICKIE ELECTRONICS SECTIONQR3 – 3 Button Rehab ControlSUNRISE PART # 020012• Works with all Quickie Rehab Motor Controls.• Mod

Seite 119 - PG DRIVES TECHNOLOGY

41©2006 Sunrise MedicalQUICKIE ELECTRONICS SECTIONActuator Mode There are 6 different Actuator Modes to go through.Lights will illuminate to indicate

Seite 120 - 120©2006 Sunrise Medical

42©2006 Sunrise MedicalQUICKIE ELECTRONICS SECTIONFlying Lead ReplacementRemove the Joystick from the Joystick Mount1. Remove the four T 10 torx screw

Seite 121 - VR2 Remote Controller

43©2006 Sunrise MedicalQUICKIE ELECTRONICS SECTIONJoystick Housing ReplacementTurn off Joystick and unplug 1. Remove the Joystick from the Joystick Mo

Seite 122 - VR2 Plugs/Connectors

44©2006 Sunrise MedicalQUICKIE ELECTRONICS SECTIONMountingQR Hand Controls can be mounted on either • side of the Wheelchair to accommodate Left Hande

Seite 123 - VR2 Without Power Seating

45©2006 Sunrise MedicalQUICKIE ELECTRONICS SECTIONWiringBefore making any connections to the Hand • Control, disable the Wheelchair by disconnecting t

Seite 124 - VR2 DRIVE-THRU TILT

46©2006 Sunrise MedicalQUICKIE ELECTRONICS SECTIONDeviceType of input control utilized : hand control,3 Button, SCIM1, SCIM1 or compact joystick.Forwa

Seite 125 - R-NET Remote Controller

47©2006 Sunrise MedicalQUICKIE ELECTRONICS SECTIONPower On / Off OptionsThis button turns the wheelchair control on and off. You can see if the wheel

Seite 126

48©2006 Sunrise MedicalQUICKIE ELECTRONICS SECTIONHand Held ProgrammerHand Held ProgrammerFeatures :(+) and (-) keys (A) to increase or decrease • dat

Seite 127 - R-NET Optional Electronics

49©2006 Sunrise MedicalQUICKIE ELECTRONICS SECTIONConnector FunctionsSymbolBatteryBattery ChargerMotor/Parking Brake A,BEncoder A,BConnector Function

Seite 128 - R-NET Plugs/Connectors

5©2006 Sunrise MedicalINTRODUCTIONQuickie Rhythm FactsMid Wheel Drive (MWD)• Standard 6.5 mph motor ( 300 lb. capacity) or • optional 5.0 mph motor (4

Seite 129

50©2006 Sunrise MedicalQUICKIE ELECTRONICS SECTIONWiring:Before making any connections to the Controller, disable the Wheelchair by disconnecting the

Seite 130 - Main Wiring Diagram R-NET

51©2006 Sunrise MedicalQUICKIE ELECTRONICS SECTIONCan Bus Pin OutsPin Defi nition1 CAN Hi2 CAN Lo3 +24V4 GroundElectronic Pin Out Characteristics – QC

Seite 131 - Main Wiring Diagram R-net

52©2006 Sunrise MedicalQUICKIE ELECTRONICS SECTIONCan Bus Pin OutsPin Defi nition1 CAN Hi2 CAN Lo3 +24V4 GroundElectronic Pin Out Characteristics – QR

Seite 132 - 2009SUNRISE MEDICAL

53©2006 Sunrise MedicalQUICKIE ELECTRONICS SECTIONThe fi rst step in Troubleshooting any Power Wheelchair is to look at the chair and break it down int

Seite 133 - Section 1

54©2006 Sunrise MedicalQUICKIE ELECTRONICS SECTIONLED illumination state Error Description Priority * Actuator Internal Error 2 Module Communicatio

Seite 134

55©2006 Sunrise MedicalQUICKIE ELECTRONICS SECTIONDelphi Diagnostics Trouble Shooting ChartSymptom Code Flash Diagnosis SolutionBattery Voltage has Dr

Seite 135 - Section 2

56©2006 Sunrise MedicalQUICKIE ELECTRONICS SECTIONIntroductionObserve all safety precautions in the Quickie iQ, 1. the HHP Owner’s Manual and the Rhyt

Seite 136

57©2006 Sunrise MedicalQUICKIE ELECTRONICS SECTIONBattery and Battery Charger TestsNotes:Fully charge the batteries prior to testing. Fully • charged

Seite 137

58©2006 Sunrise MedicalQUICKIE ELECTRONICS SECTIONOperate the chair for one minute at high speed. (Gear boxes disengaged). Observe and record the vo

Seite 138

59©2006 Sunrise MedicalQUICKIE ELECTRONICS SECTIONThe battery voltage should slowly increase from its starting value once the charger is connected. •

Seite 139 - Section 3

6©2006 Sunrise MedicalINTRODUCTIONModule Error Battery Operating State of Charge Indicator State Indicator Seating IndicatorController Internal ErrorD

Seite 140

60©2006 Sunrise MedicalQUICKIE ELECTRONICS SECTIONand V. (fi g2.23.12) If using the Quickie iQ, select Monitor Parameter, followed by selecting Motor

Seite 141

61©2006 Sunrise MedicalQUICKIE ELECTRONICS SECTION( fi g 2.23.14) If using the Quickie iQ, select Monitor Parameter 10. followed by selecting Brakes. (

Seite 142

62©2006 Sunrise MedicalQUICKIE ELECTRONICS SECTIONGear Box Test Perform the following steps to electrically test the gear box. Prior to completing th

Seite 143

63©2006 Sunrise MedicalQUICKIE ELECTRONICS SECTIONIf using the Quickie iQ, select Monitor Parameter followed by selecting Hand Control, then JS Positi

Seite 144 - SUNRISE MEDICAL

64©2006 Sunrise MedicalQUICKIE ELECTRONICS SECTIONOperate each button (or toggle switch) one at a time to verify that they function correctly. This i

Seite 145 - Section 4

65©2006 Sunrise MedicalQUICKIE ELECTRONICS SECTIONIf using the Quickie iQ, select Monitor Parameter 4. followed by selecting SCM1 or SCM2, then Switch

Seite 146

66©2006 Sunrise MedicalQUICKIE ELECTRONICS SECTIONActuator Test (Motor Controller controlled actua-tors)Complete the following steps to test the seati

Seite 147

67©2006 Sunrise MedicalQUICKIE ELECTRONICS SECTIONIf using Quickie iQ, this value may be found under Programmed Parameters Access, Motor Control, • th

Seite 148 - PAGE 4.4

68©2006 Sunrise MedicalQUICKIE ELECTRONICS SECTIONActuator Test (QMAC controlled actuators)Complete the following steps to test the seating system act

Seite 149 - Module Error

69©2006 Sunrise MedicalQUICKIE ELECTRONICS SECTIONIf using the HHP, this value may be found under • Program, then QMAC, and then select the ac-tuator

Seite 150 - Section 5

7©2006 Sunrise MedicalINTRODUCTION7©2002 Sunrise MedicalBasic Setup When setting up the components of the chair, complete the following checklist to e

Seite 151

70©2006 Sunrise MedicalQUICKIE ELECTRONICS SECTIONGeneral Wheelchair Test Complete the following test to verify Time On, Time Driven, and Internal Con

Seite 152

71©2006 Sunrise MedicalSEATINGPerfect Fit Seating SystemThe integrated Perfect Fit Seating System is available in both non-recline, and recline with s

Seite 153

72©2006 Sunrise MedicalSEATING SECTIONThe Posterior Lateral Positioning Back contains:Deeply curved lateral wings to provide a lateral • hugHeadrest a

Seite 154

73©2006 Sunrise MedicalSEATING SECTIONThe Power Recline with Shear Reduction provides:Shear induced during full angle range: 2” to • 2.5”Back reclines

Seite 155

74©2006 Sunrise MedicalSEATING SECTIONRefer to the following tables for the range of adjustments for the Perfect Fit Seating System.Narrow Center PanM

Seite 156

75©2006 Sunrise MedicalSEATING SECTIONBoth the seat back and the seat pan width adjust independently. To adjust either the seat back or seat pan: Loc

Seite 157 - Section 6

76©2006 Sunrise MedicalSEATING SECTIONNote: Each starting back height adjusts upward 3”. The adjustment is based on a slot & block alignment. Ref

Seite 158

77©2006 Sunrise MedicalSEATING SECTIONThe seat has 3” to 4” of built in depth adjustment depending on the starting depth. Note: Seat Depth Adjustment

Seite 159

78©2006 Sunrise MedicalSEATING SECTIONTwo versions of the manual back adjustment are available, the Adjustable Integrated Back and the Manual Recline

Seite 160

79©2006 Sunrise MedicalSEATING SECTION1. The Perfect Fit Flip Back Height Adjustable Armrest is only available with the manually reclining backrest.2.

Seite 161

8©2006 Sunrise MedicalINTRODUCTIONMultimeter TutorialThe multimeter is one of the most useful tools in the toolbox. It can be used to check wires, sho

Seite 162 - PAGE 6.6

80©2006 Sunrise MedicalSEATING SECTIONLevel the Arm PadComplete the following steps to level the arm pad:Loosen the upper clamps.• Pull the armrest ou

Seite 163 - Section 7

81©2006 Sunrise MedicalSEATING SECTIONComplete the following procedures to adjust the angle of the armrest. Loosen the outer clamp. (A)1. Turn the re

Seite 164

82©2006 Sunrise MedicalSEATING SECTIONMount and Adjust Lateral Knee Supports using • the two mounting screws (Fig 3.37 & Fig 3.38). To open Latera

Seite 165 - Section 8

83©2006 Sunrise MedicalSEATING SECTIONNew leg rest hangers are used on the Perfect Fit Seating System. These hangers provide more positive lock. The

Seite 166

84©2006 Sunrise MedicalSEATING SECTIONLateral supports come in several different styles, • mounts and offsets. Mount them on the seat back wings as s

Seite 167

85©2006 Sunrise MedicalSEATING SECTIONHead rests come in different styles. Several • third party head rests such as Whitmyer, Otto Bock and Jay may be

Seite 168

86©2006 Sunrise MedicalSEATING SECTIONTheory of OperationThe Recline Portion of the Perfect Fit Seating System uses dual keipers to operate the back.

Seite 169

87©2006 Sunrise MedicalSEATING SECTIONThe PC board has several functions, including routing the voltage from the controller module to the recline actu

Seite 170

88©2006 Sunrise MedicalSEATING SECTIONPrior to troubleshooting the Perfect Fit seating system, verify that the QR-MAC is correctly programmed and that

Seite 171

89©2006 Sunrise MedicalSEATING SECTIONIf the readings are not in the acceptable range 9. (3 to 9 ohms), replace the keiper assembly and retest. Locat

Seite 172

9©2006 Sunrise MedicalINTRODUCTIONThis section describes the basic symbols used in a typical multimeter.ACThis symbol means alternating current. Use t

Seite 173

90©2006 Sunrise MedicalSEATING SECTIONThe chair will not drive or will only drive in creep modeVerify that the battery charger is not plugged into 1.

Seite 174

91©2006 Sunrise MedicalSEATING SECTIONTilt or Lift, check Encoder Reset:7. Disconnect the 6-pin actuator connector (A) a. from the control module.Conn

Seite 175

92©2006 Sunrise MedicalSEATING SECTIONIf any reading is not correct, disconnect each 4. cable at the actuator and check for continuity through it to t

Seite 176

93©2006 Sunrise MedicalSEATING SECTIONVerify that the BUS cables leading to the seating 3. system are mated at the back of the wheelchair. If disconn

Seite 177

94©2006 Sunrise MedicalSEATING SECTIONDisconnect the 2-pin connector and measure for 7. battery voltage on the cable leading to the base. If battery v

Seite 178

95©2006 Sunrise MedicalSEATING SECTIONIndividual Actuators Do Not Function or Function CorrectlyVerify that the seating functions illuminate on the in

Seite 179 - Section 9

96©2006 Sunrise MedicalSEATING SECTIONActuator Approximate Reading In Mid-StrokeApproximate Reading at Home PositionLeft Leg (for powered foot platfor

Seite 180

97©2006 Sunrise MedicalSEATING SECTION Motor Armature – approximately 5 ohms. Switch pressed – Open Switch not pressed – Closed Switch pressed – Clos

Seite 181

98©2006 Sunrise MedicalSEATING SECTIONWith the back pan set at 95° ±1° with respect 1. to the seat pan, the Actuator Magnet (B) should activate Reed S

Seite 182 - 182©2006 Sunrise Medical

99©2006 Sunrise MedicalSEATING SECTION Tilt System50 degrees of tilt. • Linak Actuator. • Power Modules112233665544Pin 1 Motor + BrownPin 2 Encoder Re